Compliance UK

Washer-disinfector daily checks

One of the 4 decontamination checks in Compliance UK's dental pack, with the fields to record, who usually does it and where it sits in the CQC framework.

What this check is

By default, Compliance UK schedules it every day. You can change the schedule. The dental nurse usually does it.

The record is a signed log. Whoever does the check signs each entry. A signed entry can't be edited or deleted afterwards.

Each washer-disinfector in your equipment register gets its own copy of this check and a QR label that opens its log on a phone.

Area
Decontamination (HTM 01-05)
Default schedule
Daily
Usually done by
Dental nurse
Evidence kept
Signed log
Equipment
Washer-disinfector

What to record each time

The log for this check has 3 fields, each with its valid range beside it where there is one.

Log fields for washer-disinfector daily checks
FieldTypeUnitValid range
Spray arms rotate freelyPass or failNoneNot applicable
Door seals intactPass or failNoneNot applicable
Cycle completed without faultPass or failNoneNot applicable

On the phone, after scanning the QR label

Spray arms rotate freely
Pass or fail PassFail
Door seals intact
Pass or fail PassFail
Cycle completed without fault
Pass or fail PassFail

Where it sits in the CQC framework

This check is evidence for one CQC quality statement.

Infection prevention and control · Safe

We assess and manage the risk of infection. We detect and control the risk of it spreading and share any concerns with appropriate agencies promptly.

It also maps to Regulation 12: Safe care and treatment of the Health and Social Care Act 2008 (Regulated Activities) Regulations 2014.

Other decontamination checks

Official sources