HTM 01-05 decontamination checks

HTM 01-05 is the Department of Health and Social Care's guidance on decontamination in primary care dental practices. Compliance UK's dental pack schedules 4 decontamination checks. Here is what each one records.

Autoclave daily tests log

Default schedule
Daily
Usually done by
Dental nurse
Evidence kept
Signed log
Log fields for Autoclave daily tests log
FieldTypeUnitValid range
Automatic control test passedPass or failNoneNot applicable
Steam penetration (Helix / Bowie-Dick) passedPass or failNoneNot applicable
Holding temperatureReading°C134 to 137°C
Holding pressureReadingbar2.1 to 2.3 bar
Holding timeReadingmin3 to 4 min

More about this check

Washer-disinfector daily checks

Default schedule
Daily
Usually done by
Dental nurse
Evidence kept
Signed log
Log fields for Washer-disinfector daily checks
FieldTypeUnitValid range
Spray arms rotate freelyPass or failNoneNot applicable
Door seals intactPass or failNoneNot applicable
Cycle completed without faultPass or failNoneNot applicable

More about this check

Ultrasonic bath foil and soil tests

Default schedule
Every 3 months
Usually done by
Dental nurse
Evidence kept
Checklist
Log fields for Ultrasonic bath foil and soil tests
FieldTypeUnitValid range
Foil ablation test passedPass or failNoneNot applicable
Soil test passedPass or failNoneNot applicable

More about this check

Decontamination equipment validation

Default schedule
Every year
Usually done by
Practice manager
Evidence kept
Document or certificate

More about this check

Read HTM 01-05